Before testing, all reagents were returned to room temperature. The
test should be carried out at room temperature.
I. sample extraction (see figure 1)
Add 400 L(~10 drops) of sample buffer vertically into the sample
extraction tube, then insert the sample into the tube and rotate it
against the inner wall for about 10 times, so that the sample can
be dissolved in the solution as much as possible.
2. Sampling along the inner wall of the extraction tube, leaving as
much liquid as possible in the tube, then taking it out and
discarding it.
3. Cover the dropper.
Ii test procedure (see figure 2)
1. Take out the test card from the sealed bag.
2. Add 2 drops (~80 L) of processed sample extract to the sample
well of the test card, and start the timer.
3. Place the test card at room temperature for 15 minutes and read
the results. The result is invalid after 20 minutes.
Schematic diagram of card result judgment:
Invalid (1) Result: The quality control line (line C) has no
reaction line, and the test is invalid, so the test is repeated.
(2) Negative result: red ribbon, coloring of quality control line
(line C).
(3) Positive results: two red bands, the detection line (T line)
and the quality control line (C line) are colored.
![]() |