Brand Name : | NCoV Antigen Detection(Immunochromatography) Sample Requirements |
Model Number : | 2019-nCoV Antigen(Immunochromatography) |
Certification : | CE BfArm |
Price : | To be negotiated |
Supply Ability : | 10000 Kit/Kit per Day |
Delivery Time : | To be negotiated |
[Test Method] Please restore all reagents to room temperature before testing, and the testing shall be conducted at room temperature.
Extraction of specimens (see Figure 1)
1. adds 400 μL(about 10 drops) sample buffer vertically to the
sample extraction tube, then inserts the sample test into the
sample extraction tube, rotating closely against the inner wall
about 10 times to make the specimen dissolve in the solution as
much as possible.
2. shall squeeze the cotton swab head of the test tube along the
inner wall of the extraction tube to keep the liquid in the tube as
much as possible, and remove and discard the test son.
3. Cover with the drops.
Test procedure (see Figure II)
The 1. removes the detection card from the sealing bag.
2. processes the sample extract with 2 drops (about 80 μL) to the
sample hole of the test card, and then activates the timer.
Read the results when the 3. test card is placed at room
temperature for 15 minutes. Invalid read result after 20 minutes
Judgment figure of the test card result shows:
Invalid ① result: the quality control line (C line) has no reaction
line, invalid detection, and should be retested.
② negative result: a red ribbon, quality control line (C line)
color display.
③ positive results: two red bands, test lines (T) and mass control
lines (C).
![]() |